Plant Growth Chamber Standard
The purpose of Plant Growth Chamber testing is to provide evidence on the how the quality of drug substances or drug product varies with time under the influence of environmental factors such as temperature, humidity and light and to establish a re test period for the drug substance or product under recommended storage conditions. Hence Plant growth chamber serves as an important quality attribute of the product.
Hamco make GMP Plant Growth Chamber test chambers are especially built to pass FDA and ICH Q1A compliance standards and have a reputation for reliability, quality and performance. Exceed FDA/ICH guidelines for long term, intermediate and accelerated Plant growth studies.
- Temperature range 20.0 °c to 60.0 °c. Accuracy +/- 0.2 °c. Uniformity +/- 1.0 °c
- Humidity range: 40% to 95% rh, accuracy +/- 3% Rh, uniformity +/- 3% rh.
- Temperature & humidity display: HMI touch screen 4.3” with colour display.
- Heating system: u shaped fin type Nichrome wire heater in ss sheathing.
- Cooling system: hermetically sealed CFC free Emerson compressor with r 134a refrigerant.
- System is made up of split type condenser unit in well-ventilated area to avoid throw of hot air in the room
- Non condensing steam injection type humidity system with standby heater & auto switch overs.
- Stand by humidity system
- Air circulation: flange motor with impeller and blower for forced air circulation.
- Interior illumination for working area.
- Observation window – 300 x 300 mm
- Stainless steel trays
- Double walled, gap between walls filled with Puf insulation.
- Double walled with 80 mm insulated modular with panels with easy on-site assembly
- Standard model: stainless steel mirror polished chamber and heavy gauge Powder- Coated CRC exterior.
- GMP model: stainless steel mirror polished chamber and exterior of dull finish stainless steel.
- Stainless steel trays. Adjustable tray height adjustment.
- HMI touch screen controller of 4.3”/ 7” with data logging facility.
- Data logging of at least 1000 data is possible.
- Real time graph visualisation facility.
- Multiuser with password protection facility.
- High speed Ethernet based communication.
- Audio/visual alarm for deviation.
- All operations through a single 4.3”/7” colour touch screen
- Over heat temperature and humidity cut off with alarm.
- Low temperature and humidity alarm.
- Low boiler water level safeguard.
- Electrical overload compressor cut off.
- Time delay for compressor switch on.
- Distilled water or DM Water Line. ( consumption will be 20 ltrs/24 hrs upto 450 ltr Model )
- Room temperature around machine preferably at 25 degree C.
- Stabilised input voltage of 230V Ac, 50 Hz, 20 Amps.
- Water drain line with ½” Nozzle to be provided at floor level.
- 21 CFR part 11 compliant software with audit trial ( who did what and when is recorded)
- Multilevel passwords.
- Data storage device has online and offline capabilities.
- Tabular and graphical report, alarm deviation report.
- Mean kinetic temperature report and alarm deviation report.
- Audit trial report.
- Attractive reports and graphs.
- Alarm report.
- Time synchronisation.
Ordering Information:
MODEL NO. | CAPACITY | DIMENSION IN CMS | Tray Qty. | Tray size in mm | |||
---|---|---|---|---|---|---|---|
Standard | GMP | Cu. Ft | Litres | Chamber size in mtr (h xw xd) | Overall size in mtr (hxwxd) | ||
Hamco63s -190 | Hamco63g-190 | 190 | 5400 | 2.01.362.0 | 2.161.522.16 | 24 | 870300 |
Hamco63s -230 | Hamco63 g-230 | 230 | 6400 | 2.41.362.0 | 2.561.522.16 | 32 | 870300 |
Hamco63s -280 | Hamco63 g-280 | 280 | 8000 | 2.02.0×2.0 | 2.162.162.16 | 24 | 870600 |
Hamco63s -340 | Hamco63 g-340 | 340 | 9600 | 2.42.0×2.0 | 2.562.162.16 | 32 | 870600 |
Hamco63s -428 | Hamco63 g-428 | 428 | 12000 | 2.42.0×2.5 | 2.562.162.66 | 48 | 725600 |
Hamco63s -530 | Hamco63 g-530 | 530 | 15000 | 2.42.52.5 | 2.562.662.66 | 48 | 725800 |
Hamco63s -640 | Hamco63 g-640 | 640 | 18000 | 2.42.53.0 | 2.562.663.16 | 48 | 900800 |
Hamco63s -1070 | Hamco63g-1070 | 1070 | 30000 | 2.42.55.0 | 2.562.665.16 | 80 | 920800 |
Hamco63s -1785 | Hamco63g-1785 | 1785 | 50000 | 3.04.24.0 | 3.164.364.16 | 140 | 870600 |
Hamco63s -2857 | Hamco63g-2857 | 2857 | 80000 | 3.04.85.56 | 3.164.965.72 | 180 | 870600 |